EMDR Bilateral Stimulation Research Platform Features

Last Updated: November 11, 2025 Project Version: Research Platform v1.0 Hardware: Dual Seeed XIAO ESP32-C6 Framework: ESP-IDF v5.5.0 via PlatformIO


Overview

This document describes the research platform capabilities of the EMDR bilateral stimulation device, which extends beyond standard therapeutic parameters to enable scientific investigation of bilateral stimulation mechanisms.

Dual Purpose: 1. Clinical Tool: Standard EMDR therapy (0.5-2 Hz, fixed patterns) 2. Research Platform: Extended parameters (0.25-2 Hz, multiple patterns)


Research Platform Architecture

BLE Service Structure

Configuration Service (Mobile App Control) - UUID: 6E400002-B5A3-F393-E0A9-E50E24DCCA9E - Purpose: User interface for parameter configuration

Bilateral Control Service (Device-to-Device) - UUID: 6E400001-B5A3-F393-E0A9-E50E24DCCA9E - Purpose: Real-time bilateral coordination


Stimulation Patterns

1. BILATERAL_FIXED (Standard EMDR)

Traditional bilateral stimulation with fixed motor directions:

Time (ms)    Server Device    Client Device
0-250        FORWARD ON       OFF
250-500      OFF             REVERSE ON
500-750      FORWARD ON       OFF
750-1000     OFF             REVERSE ON

Applications: - Standard EMDR therapy - Baseline measurements - Control condition

Research Questions: - Optimal frequency for different conditions - Relationship between frequency and processing speed

2. BILATERAL_ALTERNATING (Research Mode)

Both devices alternate motor direction each cycle:

Time (ms)    Server Device        Client Device
0-250        FORWARD ON          OFF
250-500      OFF                REVERSE ON
500-750      REVERSE ON         OFF
750-1000     OFF               FORWARD ON
1000-1250    FORWARD ON         OFF
1250-1500    OFF               REVERSE ON

Applications: - Direction preference studies - Habituation reduction research - Motor adaptation studies

Research Questions: - Does alternating direction improve outcomes? - Impact on sensory habituation - User preference for direction patterns

3. UNILATERAL (Control Studies)

Single device operation for research controls:

Time (ms)    Server Device    Client Device
0-250        ON              OFF (always)
250-500      OFF             OFF (always)
500-750      ON              OFF (always)
750-1000     OFF             OFF (always)

Applications: - Control conditions - Baseline comparisons - Single-sided stimulation studies

Research Questions: - Bilateral vs unilateral effectiveness - Necessity of bilateral alternation - Hemispheric processing differences


Extended Parameter Ranges

Frequency Range: 0.25-2 Hz

Frequency Cycle Time Classification Research Application
0.25 Hz 4000ms Ultra-slow Dissociative disorders, slow processing
0.5 Hz 2000ms Slow Standard EMDR minimum, elderly populations
0.75 Hz 1333ms Moderate-slow Attention disorders
1.0 Hz 1000ms Standard Typical therapeutic rate
1.25 Hz 800ms Moderate-fast Enhanced processing
1.5 Hz 667ms Fast Anxiety disorders
1.75 Hz 571ms Very fast ADHD applications
2.0 Hz 500ms Ultra-fast Standard EMDR maximum

Motor Intensity: 30-80% PWM

Safety Constraints: - Minimum 30%: Ensures perceptible stimulation - Maximum 80%: Prevents motor damage and tissue irritation

Research Applications: - Sensory threshold studies (30-40%) - Standard therapy (40-60%) - Enhanced stimulation (60-80%)

Duty Cycle: 10-50%

Duty Cycle On Time (1Hz) Application
10% 50ms Micro-stimulation research
15% 75ms Minimal effective dose studies
25% 125ms Standard therapy (battery efficient)
35% 175ms Enhanced stimulation
50% 250ms Moderate duty (battery efficient)
75% 375ms High intensity stimulation
100% 500ms Maximum duty (uses entire ACTIVE period, respects INACTIVE period)
nNote: All duty levels guarantee motor is OFF for at least 50% of total cycle time. Duty cycle is percentage of ACTIVE period only.

Session Duration: 20-90 minutes

  • 20 minutes: Minimum research protocol
  • 30 minutes: Short therapy session
  • 45 minutes: Standard therapy session
  • 60 minutes: Extended therapy session
  • 90 minutes: Maximum research session

Research Protocol Examples

Protocol 1: Direction Preference Study

Objective: Determine if motor direction affects therapeutic outcome

Design: Within-subjects, counterbalanced

Procedure:

1. Baseline Assessment (5 min)
   - BILATERAL_FIXED at 1 Hz, 40% PWM

2. Condition A (10 min)
   - BILATERAL_FIXED pattern
   - Record subjective ratings

3. Rest Period (5 min)
   - No stimulation

4. Condition B (10 min)
   - BILATERAL_ALTERNATING pattern
   - Record subjective ratings

5. Post Assessment (5 min)
   - Compare conditions

Protocol 2: Minimal Stimulation Threshold

Objective: Determine minimum effective stimulation parameters

Design: Ascending method of limits

Procedure:

1. Start: 10% duty cycle, 30% PWM, 1 Hz
2. Increment: +5% duty cycle every 5 minutes
3. Measure: Subjective perception threshold
4. Continue: Until therapeutic effect reported
5. Verify: Repeat at identified threshold

Protocol 3: Frequency Optimization

Objective: Find optimal frequency for specific populations

Design: Randomized blocks

Procedure:

Day 1: 0.5 Hz (20 min)
Day 2: 1.0 Hz (20 min)
Day 3: 1.5 Hz (20 min)
Day 4: 2.0 Hz (20 min)
Day 5: 0.25 Hz (20 min) - research extension

Measure: Processing speed, subjective comfort, therapeutic outcome


Data Collection

Real-Time Parameters

Logged every stimulation cycle:

typedef struct {
    uint32_t timestamp_ms;        // Millisecond timestamp
    uint16_t commanded_cycle_ms;  // Target cycle time
    uint16_t actual_cycle_ms;     // Measured cycle time
    uint8_t motor_intensity;      // PWM percentage (30-80)
    uint8_t duty_cycle;           // On-time percentage (10-50)
    uint8_t pattern_type;         // FIXED/ALTERNATING/UNILATERAL
    uint8_t device_role;          // SERVER/CLIENT
    uint16_t sequence_number;     // For packet loss detection
    uint8_t battery_percent;      // Battery level
    int16_t back_emf_mv;         // Motor back-EMF (if enabled)
} research_data_t;

Session Metadata

Stored at session start/end:

typedef struct {
    uint32_t session_id;          // Unique session identifier
    uint32_t start_time;          // Unix timestamp
    uint32_t duration_ms;         // Total session length
    uint8_t protocol_version;     // Research protocol ID
    uint16_t total_cycles;        // Number of bilateral cycles
    uint16_t packet_loss_count;   // BLE reliability metric
    uint8_t pattern_changes;      // Number of pattern switches
} session_metadata_t;

Safety Features

Hard Limits (Cannot Override)

  • PWM Range: 30-80% (enforced in firmware)
  • Non-overlapping: Devices never stimulate simultaneously
  • Emergency Stop: 5-second button hold always active
  • Watchdog Timer: 2-second timeout prevents lockups

Soft Limits (User Warnings)

  • Session > 60 min: "Extended session - confirm to continue"
  • Duty > 40%: "High duty cycle may cause motor heating"
  • Frequency < 0.5 Hz: "Below standard EMDR range"
  • PWM > 70%: "High intensity - monitor for comfort"

Research Mode Indicators

LED patterns indicate research vs clinical mode:

  • Clinical Mode: Solid green during operation
  • Research Mode: Pulsing blue during operation
  • Parameter Change: Brief yellow flash
  • Safety Limit: Red flash with warning

BLE Characteristic Reference

Bilateral Control Service Characteristics

Characteristic UUID Suffix Type Range Access
Bilateral Command 0101 uint8 0-6 Write
Total Cycle Time 0201 uint16 500-4000ms R/W
Motor Intensity 0301 uint8 30-80% R/W
Stimulation Pattern 0401 uint8 0-2 R/W
Device Role 0501 uint8 0-2 Read
Session Duration 0601 uint32 20-90 min R/W
Sequence Number 0701 uint16 0-65535 Read
Emergency Shutdown 0801 uint8 1 Write
Duty Cycle 0901 uint8 10-50% R/W

Implementation Status

Completed Features

  • ✅ Basic bilateral coordination (BILATERAL_FIXED)
  • ✅ BLE Configuration Service
  • ✅ Standard EMDR frequencies (0.5-2 Hz)
  • ✅ Motor PWM control (30-90%)
  • ✅ Battery monitoring
  • ✅ Session timing

In Development

  • 🚧 BILATERAL_ALTERNATING pattern
  • 🚧 Extended frequency range (0.25 Hz)
  • 🚧 Research data logging
  • 🚧 Bilateral Control Service

Future Features

  • ⏳ Real-time data streaming
  • ⏳ Physiological sensor integration
  • ⏳ Machine learning optimization
  • ⏳ Cloud data synchronization

Comparison with Clinical Standards

Parameter EMDR Standard Research Platform Rationale
Frequency 0.5-2 Hz 0.25-2 Hz Explore slow processing
Pattern Fixed bilateral Fixed/Alternating/Uni Study mechanisms
Intensity Not specified 30-80% PWM Safety + variability
Duty Cycle Not specified 10-50% Energy + comfort
Session 60-90 min typical 20-90 min Research flexibility

Research Ethics Considerations

Required for Human Studies

  1. IRB approval for research protocols
  2. Informed consent with research disclosure
  3. Data anonymization (no PII stored)
  4. Right to withdraw from research
  5. Clear clinical vs research mode indication

Device Labeling

  • "Research features are investigational"
  • "Standard EMDR mode for clinical use"
  • "Extended parameters for research only"

Future Research Directions

Phase 2: Sensor Integration

  • Heart rate variability (HRV)
  • Galvanic skin response (GSR)
  • EEG correlates
  • Eye tracking

Phase 3: Closed-Loop Control

  • Adaptive frequency based on HRV
  • Intensity modulation via GSR
  • Pattern switching on habituation
  • Personalized protocols

Phase 4: Multi-Site Studies

  • Standardized research protocols
  • Cloud data aggregation
  • Cross-population comparisons
  • Efficacy meta-analysis

Phase 5: AI Optimization

  • Machine learning parameter selection
  • Predictive therapeutic outcomes
  • Personalized treatment protocols
  • Real-time adaptation algorithms

References

  1. Shapiro, F. (2001). Eye Movement Desensitization and Reprocessing (EMDR): Basic Principles, Protocols, and Procedures.

  2. EMDRIA Standards: https://www.emdria.org

  3. ESP-IDF Documentation: https://docs.espressif.com/projects/esp-idf/

  4. NimBLE Documentation: https://mynewt.apache.org/latest/network/


Contact & Contributing

Repository: https://github.com/lemonforest/mlehaptics License: GPL v3 (Software), CERN-OHL-S v2 (Hardware) Research Collaboration: Open to academic partnerships


This document describes research capabilities that extend beyond standard EMDR therapy. Research features should only be used in appropriate research contexts with proper oversight and consent.